- Other Lab Supplies[4]
- Measuring & Analysing Instrument Processing Services[10]
- Reactors[10]
- Other Measuring & Analysing Instruments[1]
- Spectrometers[7]
- Other Physical Measuring Instruments[2]
- Measuring & Analysing Instrument Stocks[6]
- Pumps[3]
- Other Machinery[1]
- Drying Equipment[9]
- Refrigeration & Heat Exchange Parts[2]
- Laboratory Thermostatic Devices[1]
- Clinical Analytical Instruments[7]
- Pipette[4]
- Ice Makers[9]
- Sterilization Equipments[3]
- Chemical Product Machinery[1]
- Contact Person : Ms. DUAN KATHY
- Company Name : Shaanxi Pengzhan Technology Co., Ltd.
- Tel : 86-187-29970929
- Fax : 86-29-82224225
- Address : Shaanxi,Xi'an,Room 1010,Sheng Du Mansion,Building 3 o,No.379,Electronic Road,Yanta District,Xi'an City
- Country/Region : China
- Zip : 710065
Related Product Searches:ICP spectrometer for metal analysis(Inductively Couple Plasma Spectrometers),High Quality,icp spectrometer for metal analysis, icp test,PZ-ICP-2000+
Inductively Coupled Plasma Spectrometers (ICP)
Instrument introduction
PZ-ICP-2000 Inductively Coupled Plasma Spectrometers can detect approximately 70 trace to constant elements in different materials. They are widely used in scientific research Institutes, colleges, and other industry enterprises related to the fields such as geology, metallurgy, rare earth materials, electro-plating, cement, petroleum, chemical engineering, environmental detection, non-ferrous metals, medical science, food and agriculture.
Performance characteristics
Simultaneous measurement of multi elementsLow detection limit: 0.1-10ppb, and even lower Wide density linear range, 5-6 orders of magnitudeLess chemical interference, maximum accuracy, RSD≤2% High analysis speed,low cost
Technical specifications
RF Generator Specifications Circuit type: inductively feedback self-oscillation circuit, coaxial cable transmission, matched tuning, power feedback close-loop auto-controlWorking frequency: 40MHz±0.05% Frequency stability: <0.1%Power output: 800W-1200WPower output stability: <0.3%Intensity of electromagnetic leakage: electric filed intensity in 30cm distanceto the body: E:<2V/mHigh Frequency BoxHigh Frequency Box excites very stable plasma torch by exploiting the inductively feedback self-oscillation circuit. The circuit is transmitted by coaxial cable,matchingly tuned and implements close-loop auto-control by means of power feedback (National Patented Invention). Industrial argon can be used in this kind of generator, which not only makes the lighting the ICP torch easy butalso saves the cost of highly purified argon. Match boxThe high performance match box is composed of vacuum ceramic adjustable capacitors and gas adjustable capacitors, which allows it to reach the best match state and achieve maximum effective output on load coils. Directional couplerThe stability of power output is achieved by controlling the anode voltage of oscillation valve after the power signals are sent out of the directional coupler and get amplified by the amplifier.
Sample introduction system specifications:Inner diameter of output coils: 25 mm, 3 coils Torch: three concentric torches, outer diameter 20 mm, quartz Outer diameter of coaxial nebulizer: 6mmOuter diameter of double-pass spray chamber: 34mmSpecifications of Ar flowmeter and carrier gas pressure gauge: Plasma gas flowmeter: (100-1000)L/h (1.6-16L/min)Auxiliary gas flowmeter(10-100)L/h (0.16-1.66L/min)Carrier gas flowmeter10-100)L/h (0.16-1.66L/min)Carrier gas pressure stabilization valve (0-0.4MPa) Cooling water: Water temperature: 15-25°CFlow rate >5L/min Water pressure: >0.1MPa Resistivity>1MΩ
Technical Specifications of Optical SystemOptical path: Czerny-Turner Focal length: 1000mmGrating specification: Ion Beam Etching Holographic Grating; The etching density is 3600 grooves/mm; the etching area is 80*110 mm.(Alternative 2400 grooves/mm; etching area: 80*110mm)Reciprocal linear dispersion rate: 0.26nmResolution: ≤ 0.008nmScan wavelength range: 3600 grooves/mm scan wavelength range: 195—500 nm2400 grooves/mm scan wavelength range: 195—800 nm Minimum pace of stepper motor drive: 0.0006nmExit and entrance slit: 20μmSpeculum specifications: (78×105×16)mm Lensφ30,1:1 imagingThermostatic device for optical system: 32°C±1°C
Application fields
Petrochemical engineering
MineralsMetallurgyGeographyMedicine safetyLab research Environment monitoringFood safety
ICP spectrometer for metal analysis(Inductively Couple Plasma Spectrometers)